Comparison of in-phase and opposed-phase GRE and conventional SE MR pulse sequences in T1-weighted imaging of liver lesions.

J Comput Assist Tomogr. 1996 Nov-Dec;20(6):890-7. Unique Identifier : AIDSLINE MED/97087848
Martin J; Sentis M; Puig J; Rue M; Falco J; Donoso L; Zidan A; Unitat de Diagnostic per la Imatge D'Alta Tecnologia (UDIAT),; Consorci Hospitalari del Parc Tauli, Barcelona, Spain.


Abstract: PURPOSE: Our goal was to compare in-phase (IP) and opposed-phase (OP) GRE and conventional SE sequences in T1-weighted (T1-W) imaging of the liver and to evaluate chemical shift GRE imaging in characterizing liver/lesions for fat content. METHOD: IP and OP T1-W GRE with fast low angle shot (FLASH) technique and T1-W SE sequences were compared in 162 patients at 1.0 T. Chemical shift GRE imaging was used to characterize lesions with fat content. Two hundred sixteen lesions were analyzed in three groups of liver: (a) "normal" liver (n = 74 with 110 lesions); (b) cirrhotic liver (n = 76 with 85 lesions); and (c) fatty liver (n = 12 with 21 lesions). Liver/lesion contrast and liver/lesion contrast-to-noise ratio were assessed for lesion detectability. The percentage of signal intensity variation (SIV) between IP and OP images was used to characterize lesions for fat content. RESULTS: The OP GRE sequence had significantly higher contrast for normal and cirrhotic livers (p < 0.001), and the IP GRE sequence had significantly higher contrast and contrast-to-noise ratio for fatty liver (p < 0.001). There were no significant differences between OP, IP, and T1-W SE imaging in cirrhotic cases for contrast-to-noise ratio (p < 0.28). Chemical shift imaging detected fat in 21 lesions (9.7%, mean SIV, 191.1%) (sensitivity and specificity 100% when compared with fine needle aspiration cytology). CONCLUSION: OP GRE sequences could replace conventional SE sequences in T1-W imaging in nonfatty livers, whereas in fatty livers, T1-W SE sequences could be obviated, but both OP and IP sequences are necessary. Chemical shift imaging (OP and IP) can be used to accurately characterize lesions for fat content.
Keywords: Analysis of Variance Artifacts Comparative Study Echinococcosis, Hepatic/*DIAGNOSIS Evaluation Studies Female Human Liver/*PATHOLOGY/RADIOGRAPHY/ULTRASONOGRAPHY Liver Cirrhosis/*DIAGNOSIS Liver Neoplasms/*DIAGNOSIS Magnetic Resonance Imaging/INSTRUMENTATION/*METHODS/STATISTICS & NUMER DATA Male Middle Age Prospective Studies Tomography, X-Ray Computed JOURNAL ARTICLEKWDanalysisofvarianceartifactscomparativestudyechinococcosis,hepatic/KWDdiagnosisevaluationstudiesfemalehumanliver/KWDpathology/radiography/ultrasonographylivercirrhosis/KWDdiagnosisliverneoplasms/KWDdiagnosismagneticresonanceimaging/instrumentation/KWDmethods/statistics&numerdatamalemiddleageprospectivestudiestomography,x-raycomputedjournalarticle
970228
M9721868

Copyright © 1997 - National Library of Medicine. Reproduced under license with the National Library of Medicine, Bethesda, MD.

AEGiS is a 501(c)3, not-for-profit, tax-exempt, educational corporation. AEGiS is made possible through unrestricted funding from Boehringer Ingelheim, Bridgestone/Firestone Charitable Trust, Bristol-Myers Squibb Company, Elton John AIDS Foundation, Gill Foundation, the National Library of Medicine, Quest Diagnostics, Roche and Trimeris, and donations from users like you. Always watch for outdated information. This article first appeared in 1997. This material is designed to support, not replace, the relationship that exists between you and your doctor.

AEGiS presents published material, reprinted with permission and neither endorses nor opposes any material. All information contained on this website, including information relating to health conditions, products, and treatments, is for informational purposes only. It is often presented in summary or aggregate form. It is not meant to be a substitute for the advice provided by your own physician or other medical professionals. Always discuss treatment options with a doctor who specializes in treating HIV.

Copyright ©1980, 1997. AEGiS. All materials appearing on AEGiS are protected by copyright as a collective work or compilation under U.S. copyright and other laws and are the property of AEGiS, or the party credited as the provider of the content. .